The defective device has not been returned for investigation.Verification was done on a retention sample by inserting a sufficient lubricated insertion cord into a 37 fr double lumen tube.The lubricated insertion cord is able to move smoothly within the intubation tube, hence it is not possible to verify the reported failure from the retention sample.Investigation on the fractured of the tip inside dlt is still in progress.From the preliminary investigation result, it is suspected that user was not aware of that the tip of the main tube was inside dlt which led to user's impression that the tip extender was jammed.However, a jammed tip extender will not lead to fracture of the main tube.The ifu has a warning information about withdrawal of ascope 4 broncho.The user should keep the distal tip in the neutral and non-deflected position.The user should not operate the control lever, as this may result in injury to the patient and/or damage to the endoscope.Further, according to the ifu for ascope 4 broncho family the user should always make sure that the bending section is in straight position (neutral/non-deflected) and excessive force should not be used when withdrawing the endoscope.Ambu will keep monitoring this issue and take actions if necessary.
|