| Model Number |
101/870/070CZ |
| Medical Device Problem Code |
Misassembly by Users (3133)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Situation described in complaint description is clearly user failure.Instruction for use (change of tube section) provides sufficient information that obturator shall be removed once tracheostomy tube is inserted: "insert the tracheostomy tube in accordance with currently accepted medical or surgical techniques and remove the obturator." manufacturing device history record review was not performed - user issue.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer was taking out the product (tube) from the patient for regular replacement, and forgot to remove the obturator from the patient, which prevented the customer from performing respiratory ventilation on a temporary basis.No patient injury was reported.
|
| |
|