As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of catheter fractured/detached cannot be confirmed, no complaint sample was returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the instructions for use, which is supplied to the user with this item number, contains the following statements: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.If the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Potential complications: catheter fragmentation and catheter pinch-off.Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Pinch-off syndrome: pinch-off syndrome signs may include difficulty in aspirating blood, resistance to flushing or infusion of medications or fluids that improves with position changes, infraclavicular pain and/or swelling with catheter flushing or infusion palpitations, sudden onset chest pain, cardiac arrhythmias, extra heart sound, chest wall swelling at the port pocket, vein insertion site, pain in shoulder or port area not associated with swelling, cough, paresthesia of arm on side of catheter withdrawal occlusion or swishing sound with catheter flushing.Warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Note: if infusion or aspiration is successful upon lifting arm above the head and turning the head, consider pinch-off syndrome as a possible cause.The line should be radiologically evaluated if pinch-off syndrome is suspected.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|
|
A physician reported an issue with a ct w/8fr det poly cath w/vi smartport during a routine end of treatment port removal procedure.During the removal, and while making the purse string around the catheter tubing to close the tunnel, the physician pulled the suture through to tying length in preparation of removing the catheter.When doing so, the vicryl suture (not the needle, but the suture itself) completely transected the catheter.This was before the physician attempted to tighten it, nearly losing a portion of the catheter in the patient's vessel.In order to prevent migration of the catheter, the physician performed suction and then gently dissected, allowing for grasping of the catheter with hemostats.There was no report of the patient experiencing any adverse effects, harm, or require medical intervention as a result of this incident.
|