Catalog Number 121881752 |
Device Problem
Fracture (1260)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 06/21/2022 |
Event Type
Injury
|
Event Description
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for bowel dysfunction/sacral nerv stim.It was reported that the first implant only worked a little in the beginning however it was less than 50% improvement.Patient said that they made an adjustments however that didn't help.Patient said that later went to see hcp and was told that it wasn't working, doesn't recall any detailed reason why wasn't working however said that hcp replaced the system on (b)(6) 2021.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product complaint #(b)(4).Investigation summary: the device associated with this report was received for analysis.Visual examination of the returned device confirmed the allegation.The ceramic insert rim was fractured in it entirety.Fragments of the device were returned however a several portion of the rim was not returned.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: a manufacturing records evaluation (mre) was not performed as no lot number was provided for this device.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|