| Model Number |
198 |
| Medical Device Problem Code |
Suction Failure (4039)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is in-progress.All information reasonably known as of 30 jun 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during setup of the closed suction catheter (csc), the device was found to be on constant vacuum regardless of thumb control suction button position.The catheter was removed from use and a new working device was added to the patient's ventilator circuit.The patient tolerated the procedure well, with no harm.
|
| |
|
Additional Manufacturer Narrative
|
|
All information reasonably known as of 02 aug 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
Avanos medical inc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the first of two reports.Refer to 8030647-2022-00172 for the second report.
|
| |
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.The device history record for lot 30088694 was reviewed and the product was produced according to product specifications.All information reasonably known as of 24 aug 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
At receipt of the sample, the position of the thumb valve appeared in the upright position.The inside of the control valve was examined, and no visible damage was noticed; when the valve was pressed and released, it stayed in a slightly downward position and did not return to the full upright position; however, it could be manually pulled up into a full upright position.During testing, the device exhibited continued suctioning, due to the condition of the thumb valve; however, the thumb valve could be manually rotated to the locked position with no issue.The reported event could be confirmed as reported.The root cause is undetermined.All information reasonably known as of 06 jan 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|