| Model Number |
N/A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was discovered to have missing components.Attempts to obtain additional information have been made; however, no more is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The complaint device has been returned, but the device investigation has not yet been completed.Once the evaluation is completed, a supplemental medwatch will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g4, g7, h1, h2, h3, h6, h10.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows signs of repeated use and has one of components 1 and 2 disassembled / missing.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|