• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEROYAL INDUSTRIES, INC. CMC AV FISTULA PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEROYAL INDUSTRIES, INC. CMC AV FISTULA PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Model Number 89-10536
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
Deroyal cmc av fistula pack contains 4-0 silk ties instead of 2-0 silk ties as stated on the packing list.
 
Additional Manufacturer Narrative
A user facility medwatch report (b)(4).Was received on 06/23/2022.Reporting deroyal cmc av fistula pack contains 4-0 silk ties instead of 2-0 silk ties as stated on the packing list.The sample of the 4-0 silk ties were not returned to deroyal for evaluation.Root cause: determined to be a pulling and assembly slip lapse error.Corrective action: all employees responsible for this pulling / assembly error has been retrained on, acd.Qlp.034 work order process, on specific section regarding checking components to catalog numbers for accuracy.An inventory check was made by deroyal, a total of 180 of the 2-0 silk sutures 55-3088 were inspected, and no discrepancies were identified during the inspection.Also, verified 4 components / lot numbers on batch 90480 for accuracy and correct components for the employee that made the pulling error.This investigation is complete at this time.No further information is available at this time.We will provide follow-up report if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMC AV FISTULA PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
DEROYAL INDUSTRIES, INC.
1501 east central avenue
lafollette TN 37766
Manufacturer (Section G)
DEROYAL INDUSTRIES, INC.
1501 east central avenue
lafollette TN 37766
Manufacturer Contact
melissa logsdon
200 debusk lane
powell, TN 37849
8653626157
MDR Report Key14930581
Report Number3005011024-2022-00026
Device Sequence Number1970165
Product Code LRO
UDI-Device Identifier00749756364003
UDI-Public00749756364003
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K842648
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number89-10536
Device Catalogue Number89-10536
Device Lot Number56618660
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/23/2022
Initial Report FDA Received Date07/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-