• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH ARTIS ZEE BIPLANE; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH ARTIS ZEE BIPLANE; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10094141
Medical Device Problem Codes Output Problem (3005); Radiation Output Failure (4027)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.Internal id # (b)(4).
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the artis zee biplane system.During an interventional procedure.The x-ray aborted due to plane b being out-of-order.The procedure was successfully completed on an alternate system.There is no report of impact to the state of health of any patient or user involved.Siemens has requested additional information in order to conduct an investigation of the reported event.The reported event occurred in japan.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The investigation was performed considering complaint description, cs reports, system history, and system log files.Log file investigation showed a can error and collimator problems which indicate an issue with the can cable.The "bypass" mode is a mitigation for this kind of failure scenario so that a procedure can be aborted in a controlled manner if necessary.The affected part was replaced by the regional service organization which resolved the problem.The occurrence rate of the error pattern was checked.A possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE BIPLANE
Common Device Name
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forchheim 91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1- or-
rittigfeld 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
mc 65-1a
malvern, PA 19355
4843231631
MDR Report Key14931771
Report Number3004977335-2022-32035
Device Sequence Number16641656
Product Code OWB
UDI-Device Identifier04056869010069
UDI-Public04056869010069
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181407
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Inspection,Repair
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10094141
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2022
Supplement Date Received by Manufacturer08/10/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-