• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMORAL STEM 12/14 NECK TAPER; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMORAL STEM 12/14 NECK TAPER; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the sterile package was damaged.No adverse events have been reported as a result of the malfunction and additional information on the reported event is unavailable.
 
Additional Manufacturer Narrative
(b)(4).Report source: china.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4, d9, g3, g6, h2, h3, h6, h10.The following sections were corrected: d4 visual evaluation of the returned product confirmed the sterile pouch was likely punctured by the distal tip of the stem, resulting in the pouch tearing.The device history records were reviewed and no discrepancies were identified.The condition of the device when it left zimmer biomet is conforming to specification and therefore the root cause of the reported event is attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL STEM 12/14 NECK TAPER
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14933584
Report Number0001822565-2022-01905
Device Sequence Number5022432
Product Code LPH
UDI-Device Identifier00889024131675
UDI-Public(01)00889024131675(17)301013(10)64798641
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K032726
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00771100610
Device Lot Number64798641
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2022
Supplement Date Received by Manufacturer10/24/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient Age84 YR
Patient SexFemale
Patient Weight45 KG
-
-