As a result of additional information, fields d1, d4, d10, and h6 have been updated.Added information to d8.D10.Concomitants: 13a2101 - cemex system fast genta 70g sn: (b)(6), 200-03-32 - one peg patella 32mm sn: (b)(6), 200-04-23 - cemented finned tib.Tra sz 2f/3t sn: (b)(6), 200-22-13 - cr tibial insert sz 2, 13mm sn: (b)(6), 232-02-02 - optetrak asy, cr porous femoral, sz 2, left sn: 1882105, 232-03-02 - optetrak asy, cr porous femoral, sz 2, right sn: (b)(6), 620-00-02 - platelet rich plasma kit with spray tips sn: (b)(6), 620-11-02 - accelerate conc.Sys rep by 620-12-02 sn: (b)(6) h3: investigation results - the revision reported was likely the result of prosthesis wear, osteolysis, and an insufficient bond between the tibial component and the bone which led to aseptic (non-infected) tibial loosening.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.The extent and root cause of the prosthesis wear, osteolysis, and tibial loosening could not be determined as the devices were not returned for evaluation, and images and radiographs were not provided.
|