MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARCOS 15X200MM PRX TPR DIST; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARCOS 15X200MM PRX TPR DIST; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Loosening of Implant Not Related to Bone-Ingrowth (4002); Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Foreign: country: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-01574.
 
Event or Problem Description
It was reported the patient underwent a right hip revision approximately 8 weeks post implantation due to the arcos distal slotted stem subsiding.No additional information.
 
Event or Problem Description
No additional information on the reported event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6 component code: mechanical (g04) - stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device history records identified no related deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOS 15X200MM PRX TPR DIST
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14934085
Report Number0001825034-2022-01573
Device Sequence Number5690264
Product Code KWA
UDI-Device Identifier00880304478497
UDI-Public(01)00880304478497(17)280102(10)363970
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K090757
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number11-300615
Device Lot Number363970
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2022
Supplement Date Received by Manufacturer07/27/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/28/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CAT# 11-301342 LOT# 111130 ARCOS CONE
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age74 YR
Patient SexMale
Patient Weight85 KG
-
-