Model Number TCLP-280
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Codes Capsular Contracture (1761); Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The events of infection and capsular contracture are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: swelling, pain, deformity, hardening, capsular contracture, baker grade unknown, fever, immediately hospitalized with a serious infection and bilateral swelling of the breast wall, significant breast pain.
 
Event or Problem Description
Patient representative reported that the patient experienced swelling, pain, deformity, hardening, capsular contracture, baker grade unknown, fever, immediately hospitalized with a serious infection and bilateral swelling of the breast wall, significant breast pain and "both implants were textured from information and belief, which from information and belief were also recalled in the later recall by the fda." the reported events of lumps and body aches have been deemed not related to the device.This record is for the right side.The device has been explanted and replaced.
 
Additional Manufacturer Narrative
Further investigation summary: with the information collected during the investigation, there is enough evidence to support that device lot number 2961172 was manufactured under controlled conditions in accordance with abbvie¿s procedures.Seal strength results were acceptable.Environmental monitoring results were found to be acceptable and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.One additional complaint was noted for devices sterilized under sterilization run 30024442.However, it is unlikely that the events related to the reported infection are associated to the sterilization methods utilized.The terminal sterilization cycle used by abbvie was developed and validated to achieve a high level of sterility assurance.The sterilization cycle has been validated to assure that the chance of a non-sterile device is less than one in a million.During the trend review of all infection complaints for the period of (b)(6)2020 through (b)(6)2022, no adverse trend was noted.Given the fact that the cause of the infection cannot be specifically associated to the manufacturing process, and there is not an adverse trend for this type of event no corrective action is deemed necessary at this time.
 
Event or Problem Description
Patient representative reported that the patient experienced swelling, pain, deformity, hardening, capsular contracture, baker grade unknown, fever, immediately hospitalized with a serious infection and bilateral swelling of the breast wall, significant breast pain and "both implants were textured from information and belief, which from information and belief were also recalled in the later recall by the fda." the reported events of lumps and body aches have been deemed not related to the device.This record is for the right side.The device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURE COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key14934160
Report Number9617229-2022-05046
Device Sequence Number17550684
Product Code FTR
UDI-Device Identifier10888628030688
UDI-Public10888628030688
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source Other,Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/20/2021
Device Model NumberTCLP-280
Device Catalogue NumberTCLP-280
Device Lot Number2961172
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/12/2020
Supplement Date Received by Manufacturer08/05/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/20/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-