MAUDE Adverse Event Report: MICROPORT ORTHOPEDICS INC. PROFEMUR® NECK NEUTRAL SHORT COBALT CHROME; HIP COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT ORTHOPEDICS INC. PROFEMUR® NECK NEUTRAL SHORT COBALT CHROME; HIP COMPONENT Back to Search Results
Model Number PHAC1202
Medical Device Problem Code Device Damaged Prior to Receipt by User (1570)
Health Effect - Clinical Codes Pain (1994); Metal Related Pathology (4530)
Type of Reportable Event Serious Injury
Event or Problem Description
Allegedly, patient had on or about (b)(6) 2017 a profemur total hip system implanted in his left hip.On or about (b)(6) 2022 plaintiff reported for revision surgery of his failed hip prosthesis.Surgeon recommended the revision surgery after patient presented with progressive left hip pain, elevated metal ion levels and large soft tissue collection.Patient's revision surgery was necessary because the device failed due to corrosion at the neck- stem junction of the device.On or about (b)(6) 2022 it was discovered that the device failed due to corrosion of the oblong taper of the profemur cocr modular nech where it seated in the pocket of the profemur titanium stem, which caused continuing and otherwise physical injury to patient.Patient has records related to previous complaints(mom complications and previous revision surgery) that were captured under incident group (b)(4).
 
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFEMUR® NECK NEUTRAL SHORT COBALT CHROME
Common Device Name
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key14934701
Report Number3010536692-2022-00245
Device Sequence Number5098756
Product Code JDL
UDI-Device IdentifierM684PHAC12021
UDI-PublicM684PHAC12021
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K091423
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberPHAC1202
Device Catalogue NumberPHAC1202
Device Lot Number1639851
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/14/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2022
Supplement Date Received by Manufacturer06/14/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-