Manufacturer's investigation conclusion: the reported inflow cannula misalignment was confirmed through the evaluation of the submitted images.It was reported that no better re-positioning was or is available for the patient.The patient remains ongoing on heartmate (hm) 3 left ventricular assist system (lvas), serial number (b)(6).The relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The heartmate 3 lvas instructions for use (ifu) (rev.G) is currently available.Section 5 of the ifu, "surgical procedures¿, provides instructions on how to insert the pump in the ventricle and instructs the user to take care to avoid orienting the inlet towards the interventricular septum as pump function will be compromised in the presence of inlet obstruction.The steps to adjust the orientation of the pump are also outlined in this section.No further information was provided.The manufacturer is closing the file on this event.
|