MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARTICULAR SURFACE USE WITH PLATE 3,4 SIZE YELLOW/C-H 14 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARTICULAR SURFACE USE WITH PLATE 3,4 SIZE YELLOW/C-H 14 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 06/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0001822565-2022-01988.Medical product: cruciate retaining cr-flex (gsf) femoral component porous item# 00575201506 lot# 62626650.Tibial tray fixed bearing size 4 item# 00595403702 lot# 64118046.Unknown 32mm patella item# unknown lot# unknown.Report source: australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation as it was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient was revised approximately three years and two and a half months post implantation due to instability and ligament laxity.The initial tibia, femur, and articular surface were replaced with a rotating hinge knee due to compromised collateral ligaments.Attempts to obtain additional information have been made; however, no more information is available at this time.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Medical product all poly petella standard cemented size 32 mm diameter 8.5 mm thickness item# 00597206532.Lot# 64087402.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.No product was returned, or pictures provided; therefore, visual and dimensional evaluations could not be performed.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: surgery notes: revision knee for instability.Surgeon stated knee has progressively become unstable due to an increase in patient¿s ligament laxity.Removed femur, poly, and tibia and revised to rhk due to compromised collateral ligaments.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE USE WITH PLATE 3,4 SIZE YELLOW/C-H 14 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14935568
Report Number0001822565-2022-01989
Device Sequence Number5666371
Product Code JWH
UDI-Device Identifier00889024207981
UDI-Public(01)00889024207981(17)230331(10)63980683
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K173057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberN/A
Device Catalogue Number00595203014
Device Lot Number63980683
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2022
Supplement Date Received by Manufacturer07/26/2022
08/10/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date08/09/2022
08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/04/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.; SEE H10.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight110 KG
-
-