• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES ABBOTT PRISM HBCORE ASSAY KIT; HEPATITIS B VIRUS CORE ANTIGEN (E. COLI, RECOMBINANT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES ABBOTT PRISM HBCORE ASSAY KIT; HEPATITIS B VIRUS CORE ANTIGEN (E. COLI, RECOMBINANT) Back to Search Results
Catalog Number 06E66-68
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Event or Problem Description
The customer stated that a false reactive prism hbcore result was generated.The following data was provided.(b)(6) 2022.Sample id (b)(6).Hbcore results: repeat reactive 0.89, 0.93, and 0.94 s/co.The sample tested nonreactive on the alinity s processing module which is believed to be correct.Supplemental testing was nonreactive/negative for hbsag and hbv pcr.No impact to donor or patient management was reported.
 
Additional Manufacturer Narrative
Trending review determined no adverse trend for the issue for the product.Historical complaint review determined there is normal complaint activity for the lot number.Device history record review did not show any potential non-conformances, or deviations.Labeling was reviewed and found to adequately address the issue under review.To evaluate clinical specificity of prism hbcore reagent lot 35001m700, the lot and a reference lot (39006m700) were tested in a specificity setup inclusive of replicates of the negative control and a negative human population panel.The testing passed all validity and acceptance criteria.All generated data demonstrates that the prism hbcore reagent lot 35001m700 is performing as expected.No deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABBOTT PRISM HBCORE ASSAY KIT
Common Device Name
HEPATITIS B VIRUS CORE ANTIGEN (E. COLI, RECOMBINANT)
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064
Manufacturer (Section G)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14936088
Report Number1415939-2022-00050
Device Sequence Number5035570
Product Code QHM
UDI-Device Identifier00380740062965
UDI-Public00380740062965
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/13/2022
Device Catalogue Number06E66-68
Device Lot Number35001M700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer07/21/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date07/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PRSM NEXT RFRB IN, 06A36-11, 01175-96.; PRSM NEXT RFRB IN, 06A36-11, 01175-96.
Patient SexUnknown
-
-