• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941832400
Medical Device Problem Codes Break (1069); Device-Device Incompatibility (2919); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The 99% stenosed target lesion was located in the mildly tortuous and mildly calcified right coronary artery.A 4.00 x 32mm synergy xd drug eluting stent was advanced through a 6fr guidezilla guide extension catheter for treatment.However, upon advancing, resistance was felt at the collar of the guidezilla.After several attempts on pushing through the resistance, the physician noticed that the stent delivery system (sds) was bent and pushed one more time and the sds snapped in half distal to mid from the hub, closer to the middle shaft.The device was removed by pulling out from the body and the procedure was completed with a 4.0x28mm synergy xd drug-eluting stent.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy xd mr us 4.00 x 32mm stent delivery system was returned for analysis in 2 sections.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal markerbands.The crimped stent outer diameter was measured using snap gauge and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found a shaft break at approximately 90.6cm distal to the distal strain relief, as well as multiple hypotube kinks at several locations along the hypotube shaft.A visual examination of the outer and inner lumen and mid-shaft section found no issues.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that shaft break occurred.The 99% stenosed target lesion was located in the mildly tortuous and mildly calcified right coronary artery.A 4.00 x 32mm synergy xd drug eluting stent was advanced through a 6fr guidezilla guide extension catheter for treatment.However, upon advancing, resistance was felt at the collar of the guidezilla.After several attempts on pushing through the resistance, the physician noticed that the stent delivery system (sds) was bent and pushed one more time and the sds snapped in half distal to mid from the hub, closer to the middle shaft.The device was removed by pulling out from the body and the procedure was completed with a 4.0x28mm synergy xd drug-eluting stent.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14936801
Report Number2134265-2022-06972
Device Sequence Number17322026
Product Code NIQ
UDI-Device Identifier08714729981169
UDI-Public08714729981169
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/16/2024
Device Model NumberH7493941832400
Device Lot Number0028949480
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2022
Supplement Date Received by Manufacturer08/03/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-