Model Number VACDSP
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cellulitis (1768); Fungal Infection (2419)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On 02-jun-2022, the following information was provided to kci by the patient: the patient has a doctor's appointment tomorrow due to the home health nurse reporting an alleged odor from the wound on (b)(6) 2022.On 08-jun-2022, the following information was provided to kci by the patient: the activ.A.C.¿ ion progress¿ remote therapy monitoring system remains in use, however, the physician prescribed doxycycline and fluconazole to treat skin irritation and a yeast infection allegedly from the v.A.C.® drape.On 28-jun-2022, the following information was provided to kci by the physician assistant: on 03-jun-2022, the physician ordered an oral antifungal medication to treat a periwound fungal infection and due to concern for periwound cellulitis, an oral antibiotic was also prescribed.A wound culture was not obtained.On 21-jun-2022, a device history record review for the v.A.C.® granufoam¿ dressing lot number 60331235v010 and associated v.A.C.® drape was completed.All end release testing of product and packaging met specifications.The v.A.C.® granufoam¿ dressing was not returned; therefore, a device evaluation could not be performed.
 
Additional Manufacturer Narrative
The v.A.C.® granufoam¿ dressing was not returned; therefore, a device evaluation could not be performed.Based on the information provided, it cannot be determined that the alleged fungal infection and possible cellulitis requiring medications are related to the v.A.C.® drape.A device history record review for the v.A.C.® granufoam¿ dressing and associated v.A.C.® drape met specifications.Device labeling, available in print and online, states: warnings: never leave a v.A.C.® dressing in place without active v.A.C.® therapy for more than two hours.If therapy is off for more than two hours, remove the old dressing and irrigate the wound.Either apply a new v.A.C.® dressing from an unopened sterile package and restart v.A.C.® therapy; or apply an alternate dressing, such as a wet to moist gauze, as approved during times of extreme need, by treating physician.Infected wounds: infected wounds should be monitored closely and may require more frequent dressing changes than noninfected wounds, dependent upon factors such as wound conditions, treatment goals.Refer to dressing application instructions (found in v.A.C.® dressing cartons) for details regarding dressing change frequency.As with any wound treatment, clinicians and patients / caregivers should frequently monitor the patient's wound, periwound tissue and exudate for signs of infection, worsening infection, or other complications.Some signs of infection are fever, tenderness, redness, swelling, itching, rash, increased warmth in the wound or periwound area, purulent discharge or strong odor.Infection can be serious, and can lead to complications such as pain, discomfort, fever, gangrene, toxic shock, septic shock and/or fatal injury.Some signs or complications of systemic infection are nausea, vomiting, diarrhea, headache, dizziness, fainting, sore throat with swelling of the mucus membranes, disorientation, high fever, refractory and/or orthostatic hypotension or erythroderma (a sunburn-like rash).If there are any signs of the onset of systemic infection or advancing infection at the wound site, contact the treating physician immediately to determine if v.A.C.® therapy should be discontinued.Dressing changes.Wounds being treated with the v.A.C.® therapy system should be monitored on a regular basis.In a monitored, non-infected wound, v.A.C.® dressings should be changed every 48-72 hours, but no less than 3 times a week, with frequency adjusted by the clinician as appropriate.Infected wounds must be monitored often and very closely.For these wounds, dressings may need to be changed more often than 48-72 hours; the dressing changing intervals should be based on a continuing evaluation of the wound condition and the patient's clinical presentation, rather than a fixed schedule.Precautions the v.A.C.® therapy system will not be effective in addressing complications associated with the following: -ischemia to the incision or incision area -untreated or inadequately treated infection -inadequate hemostasis of the incision -cellulitis of the incision area disclaimer: this information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a kinetic concepts, inc.Product malfunctioned, is defective or has caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V.A.C.® DRAPE
Common Device Name
OMP
Manufacturer (Section D)
KINETIC CONCEPTS, INC.
san antonio TX 78249
Manufacturer Contact
steven jackson
6203 farinon drive
san antonio, TX 78249
2102556438
MDR Report Key14936911
Report Number3009897021-2022-00119
Device Sequence Number15006399
Product Code OMP
UDI-Device Identifier00849554001107
UDI-Public0100849554001107172410311060331235V010
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K133276
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberVACDSP
Device Catalogue NumberM8275051
Device Lot Number60331235V010
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2022
Initial Report FDA Received Date07/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BUSPIRONE; MONTELUKAST
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-