• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IDEAL IMPLANT INCORPORATED IDEAL IMPLANT STRUCTURED BREAST IMPLANT; SALINE-FILLED BREAST IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IDEAL IMPLANT INCORPORATED IDEAL IMPLANT STRUCTURED BREAST IMPLANT; SALINE-FILLED BREAST IMPLANT Back to Search Results
Model Number 63501
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Alleged deflation.
 
Additional Manufacturer Narrative
No manufacturing defect or other abnormality was found on macroscopic or microscopic examination of the explanted device.No leak could be reproduced during explant analysis.The cause could not be determined.Updates made to section b5, d6b, d9, g2, g3, g6, h2, h3, h6, h10.
 
Event or Problem Description
Deflation resulting in explantation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDEAL IMPLANT STRUCTURED BREAST IMPLANT
Common Device Name
SALINE-FILLED BREAST IMPLANT
Manufacturer (Section D)
IDEAL IMPLANT INCORPORATED
14881 quorum drive
suite 925
dallas TX 75254
Manufacturer (Section G)
TRELLEBORG SEALING SOLUTIONS
2761 walnut avenue
tustin CA 92780
Manufacturer Contact
robert hamas
14881 quorum drive
suite 925
dallas, TX 75254
MDR Report Key14939841
Report Number3011491947-2022-00154
Device Sequence Number12315870
Product Code FWM
UDI-Device Identifier10851795006121
UDI-Public(01)10851795006121(241)63501(17)230113
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P120011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date02/13/2023
Device Model Number63501
Device Catalogue Number635
Device Lot Number63312
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2022
Supplement Date Received by Manufacturer08/16/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-