DEPUY MITEK LLC US RIGIDFIX PLA 2.7 MM BTB CROSS PIN KIT; SOFT-TISSUE ANCHOR, BIOABSORBABLE
|
Back to Search Results |
|
| Model Number |
210127 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the affiliate in france that the rigidfix pla 2.7 mm btb cross pin kit device got jammed in the cannula.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Reporter is a j&j employee.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: b3: the event date was unknown in the initial report and has been updated accordingly.B5: additional information received from the reporter indicated that the event occurred during a kennet jones type acruciate ligament ligamentoplasty procedure.It was also reported that the there was a 5 to 10 minute delay in the procedure and another kit was available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the device was received and evaluated.Only the trocar was returned, the sleeve was not returned.Upon visual inspection, it could be observed stretch marks all over the trocar, which is an evidence that the trocar was jammed with the sleeve.A manufacturing record evaluation was performed for the finished device 8l67775 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.The possible root cause for the jammed condition can be attributed to procedural variables, such handling of the device or product interaction during procedure.Considering that the trocar has the stretch marks, we can relate this issue to a bad axis alignment of the drill at the moment of insertion, the bending force applied and the higher spinning speed of the drill can contribute to the jamming condition of the sleeve and trocar.As per ifu: directions for a successfully hole drilling and trocar and sleeve interaction are provided.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
| |
|