• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS, INC. IMPRESS®; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS, INC. IMPRESS®; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 410038VER
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
The account alleges that during an embolization procedure, the catheter tip detached within the patient tibial branch artery feeding an av fistula.This occurred on the reintroduction of a glide wire into the 4f catheter post angiographic run-off - hand injection.The clinician was unsuccessful in retrieving the foreign body from the vessel with a vascular snare device.The clinical noted that foreign body was located within a vessel intended for embolization.The foreign body would simply assist in occluding the targeted vessel.It was not necessary to remove the foreign body.Several additional embolic products were deployed, both proximal and distal to the foreign body fragment successfully completing the intended embolization procedure.No additional patient consequences to report.
 
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The suspect device was returned for evaluation.The complaint was confirmed; however, the root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
14646 kirby drive
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC.
14646 kirby drive
houston TX 77047
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012084662
MDR Report Key14939939
Report Number3010665433-2022-00069
Device Sequence Number20082314
Product Code DQO
UDI-Device Identifier00884450012413
UDI-Public0884450012413
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K053171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Catalogue Number410038VER
Device Lot NumberE1617514
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2022
Supplement Date Received by Manufacturer08/30/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ANGLED GLIDEWIRE
Patient SexUnknown
-
-