| Model Number |
23-0021 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
Abdominal Pain (1685)
|
| Date of Event |
06/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The patient experienced severe pain on the left side of their diaphragm upon initialization of stimulation with an external pulse generator.Stimulation was repeated with a second external pulse generator and there was no discomfort.
|
| |
|
Additional Manufacturer Narrative
|
|
Although this incident did not result in serious patient harm, the manufacturer's investigation identified a defect in a defined group of printed circuit boards (pcbs) that are a component of the finished device.Serious harm could result if one of the potential failure modes occurs with one of the affected devices.The defect was introduced at a subcontractor that supplies the pcbs and a supplier corrective action has been initiated.The manufacturer has initiated a recall of affected finished devices and a separate recall report will be submitted to the fda.
|
| |
|