Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.Device analysis: the product was returned for evaluation.Visual inspection was conducted on the returned samples.Visual analysis of the returned sample revealed that one empty opened foil, one paper lid, one empty winding former, and a suture of product code: y943h were returned.The needle wasn¿t received.The suture was inspected, and the insertion end did not meet the requirement.This defect is caused because the suture was not properly inserted inside of the needle barrel hole resulting in short insertion.Based on the information currently available, the pull-out was identified during the investigation of the sample received.This product issue will be addressed through quality system.The product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that one empty opened box and nineteen packets of product code y943h were returned with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packets.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened, and no defects were detected.The swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issues related no defects were observed during evaluation.Functional test was performed, and the pull force result were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.
|