| Model Number |
DFR00V |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Visual Disturbances (2140)
|
| Date of Event |
11/03/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device record and complaint trending for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an intraocular lens (iol) was explanted from the patient's left eye due to fuzzy vision.The patient also noted that they could see the side of their lens.A dfr00v model lens, 21.0 diopter size, was implanted as the replacement iol.There was no incision enlargement, no vitrectomy, and no sutures performed.The patient was doing good post-explant.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9.Device available for evaluation? yes.Returned to manufacturer on: jul 8, 2022.Section h3.Device evaluated manufacturer? yes.Device evaluation: visual inspection under magnification revealed that the lens was received cut in half.The lens was cleaned and, no issues that could contribute to or cause visual disturbances could be identified.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue ¿hm-blurry vision and hm-visual disturbance¿ was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|