BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC Back to Search Results
Model Number 39200
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that sheath detachment occurred.A 9x60x120mm and a 9x60x75mm epic iliac stents were successfully deployed.However, during removal of the stent from the sheath, the physician noticed that the inner sheath on one was intact and still on the wire and was removed over the wire from the body.The second one was intact and was removed from the body as one piece.It was noticed after removal that it was starting to separate.There were no patient complications reported.
 
Additional Manufacturer Narrative
Initial reporter address 1: (b)(6).
 
Additional Manufacturer Narrative
E1: initial reporter address 1: (b)(6).Device evaluated by mfr.: the epic was returned to boston scientific for analysis.The outer sheath, tip, inner sheath, and the remainder of the device were checked for damage.The device batch number is a match to the complaint batch number.Visual inspection showed the inner sheath is still attached to the outer shaft.The inner sheath was gently tugged and did not separate from the device.Flattening of the inner sheath was measured at 2.2 cm and 5.4 cm from the distal tip.X-ray of the handle found no damage.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that sheath detachment occurred.A 9x60x120mm and a 9x60x75mm epic iliac stents were successfully deployed.However, during removal of the stent from the sheath, the physician noticed that the inner sheath on one was intact and still on the wire and was removed over the wire from the body.The second one was intact and was removed from the body as one piece.It was noticed after removal that it was starting to separate.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14940929
Report Number2134265-2022-07211
Device Sequence Number14989455
Product Code NIO
UDI-Device Identifier08714729805830
UDI-Public08714729805830
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2026
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0027420452
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer08/03/2022
Initial Report FDA Received Date07/06/2022
Supplement Report FDA Received Date08/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-