• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. CR TIBIAL INSERT SZ 2, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. CR TIBIAL INSERT SZ 2, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 200-22-13
Medical Device Problem Codes Naturally Worn (2988); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending evaluation.
 
Event or Problem Description
It was reported that this female patient had a revision on her left knee.Reason for the revision was not reported.
 
Additional Manufacturer Narrative
Section h10: (h3) based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the subsequent revision cannot be conclusively determined; no reason for revision was given.Section h11: the following sections have corrected information: (b5) as reported by legal that this female patient had a revision on her left knee (b)(6) 2019.Patient had bilateral knee replacement surgery (b)(6)2012.Reason for the revision was not reported.No additional information available.
 
Event or Problem Description
As reported by legal that this female patient had a revision on her left knee (b)(6) 2019.Patient had bilateral knee replacement surgery (b)(6) 2012.Reason for the revision was not reported.No additional information available.
 
Additional Manufacturer Narrative
D10.Concomitants: 13a2101 - cemex system fast genta 70g, sn: (b)(6).200-03-32 - one peg patella 32mm sn: (b)(6), sn: (b)(6).200-04-23 - cemented finned tib.Tra sz 2f/3t, sn: (b)(6), sn: (b)(6).200-72-11 - cr tibial insert sz 2, 11mm, slope ++, sn: (b)(6).232-02-02 - optetrak asy, cr porous femoral, sz 2, left, sn: (b)(6).232-03-02 - optetrak asy, cr porous femoral, sz 2, right, sn: (b)(6).620-00-02 - platelet rich plasma kit with spray tips, sn: (b)(6).620-11-02 - accelerate conc.Sys rep by 620-12-02, sn: (b)(6).
 
Event or Problem Description
As reported via legal documentation, a patient had bilateral knee replacement on (b)(6) 2012.They subsequently underwent left knee revision on (b)(6) 2019, approximately 6 years 11 months post primary procedure.The surgeon noted the polyethylene had an excessive amount of wear, and was degrading/shedding debris causing pain, swelling, osteolysis, synovitis, bone loss, and tissue damage.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Additional Manufacturer Narrative
H3: investigation results - the revision reported was likely the result of prosthesis wear and osteolysis.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.The extent and root cause of the prosthesis wear and osteolysis could not be determined as the devices were not returned for evaluation, and images and radiographs were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CR TIBIAL INSERT SZ 2, 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
miguel sosa
3523771140
MDR Report Key14943353
Report Number1038671-2022-00758
Device Sequence Number19979388
Product Code JWH
UDI-Device Identifier10885862040213
UDI-Public10885862040213
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K932776
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2019
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/07/2018
Device Catalogue Number200-22-13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer08/24/2022
07/13/2023
07/13/2023
Initial Report FDA Received Date07/07/2022
Supplement Report FDA Received Date09/15/2022
08/22/2023
01/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/08/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0019-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.; UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-