| Model Number |
RLT261416 |
| Medical Device Problem Codes |
Microbial Contamination of Device (2303); Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Aneurysm expansion (4597); Clinical Endoleak (4843)
|
| Date of Event |
12/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: gore® excluder® aaa endoprosthesis - stent graft contralateral leg endoprosthesis (b)(4).Gore® excluder® aaa endoprosthesis - stent graft aortic extender endoprosthesis (b)(4).(b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2015, a patient presented with an abdominal aortic aneurysm (measured 50.6 mm) was treated with gore® excluder® aaa endoprosthesis (b)(4).On (b)(6) 2021, the patients was admitted with type ii endoleak and a sudden growth of the aneurysmal sac (80 mm).The endoleak was embolized with some metallic spiral and glue.In was additionally reported that after the procedure the patient started with a fever and covid positive results and was moved to covid ward.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 code 3331: a review of the manufacturing records indicated the lot met all pre-release specifications.Per the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to: endoleak.This device is also reported with the manufacturer report number 3007284313-2022-02018 (for an event at a later date).Corrected h6: updated health effect - impact code to 4641 - unexpected medical intervention.Code 4625 was inadvertently entered, but correct code should be 4641.Updated investigation findings code to 213 - no device problem found.Code 3233 is no longer applicable.Updated investigation conclusions code to 4315 and 22.Code 11 is no longer applicable.
|
| |
|
Event or Problem Description
|
|
After the flogistic index continues to be high a month later, an antibiotic therapy and infective consultation was initiated.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: additional event description: after the flogistic index continues to be high a month later, an antibiotic therapy and infective consultation was initiated.Per the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to: endoleak, infection (e.G., aneurysm, device or access sites).Corrected h6: updated health effect - clinical codes to e050102 - aneurysm expansion , and e2121 - endoleaks.Updated health effect - impact code to f1901 - additional surgery.Code 4641 was inadvertently entered, but correct code should be f1901.Updated medical device problem code to a180105 - microbial contamination of device, as infective consultation was initiated due to persistent high flogistic index.Updated component code to g07001 - part/component/sub-assembly term not applicable.Code 515 was inadvertently entered, but correct code should be g07001.Removed investigation conclusions code 4315.
|
| |
|
Search Alerts/Recalls
|