The reported event is confirmed, cause unknown.Two coude tip tiemann catheters were returned unopened.The samples were visually inspected and the surface of both catheters appeared to be deteriorated from oxidation, as it is unknown when the oxidation occurred, the reported event will be confirmed cause unknown.No additional defects were noted on the catheter; the eyelets were smooth, the coude bend was observed to align with the coude indicator.The devices were returned unused; therefore, were not used on a patient but are intended for treatment purposes.As the cause of the oxidation is unknown, unable to determine whether the device had met relevant specifications.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "instructions for use indications for use: for urological use only.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Caution: this product contains natural rubber latex which may cause allergic reactions.Caution: federal (u.S.A.) laws restricts this device to sale by or on the order of a physician.Warnings: ¿ on latex and red rubber catheters, do not use ointments or lubricants having a petrolatum base.They will damage latex.¿ visually inspect the product for any imperfections or surface deterioration prior to use.¿ reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Adverse reactions urinary tract infection bleeding from the urethra irritation of the urethra instructions for intermittent catheters: 1.Wash your hands thoroughly with soap and water.2.Remove catheter from the pack.3.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.4.Gently insert rounded end of catheter into urethra.5.When urine stops flowing, remove catheter from urethra.6.Dispose of catheter in accordance with local rules and regulations.7.Wash your hands." the actual/suspected device was inspected.
|