Udi: (b)(4).Device history record reviews were completed for prime plus sn: (b)(6), microsensor card pn 61615 lot: 22063005, calibrator pack pn 57831 lot: 22047040, qc pack pn 57841 lot: 21200036 and reference sensor pn 57823 lot: 22061045.The dhrs confirmed that all products met all specifications when released to finished goods.A 2 year look back in the data base did not return any result for lot/sn in question.Per the investigation report, the discrepancies in patient sample are believed to have resulted from reference flow issue which can be caused by debris entering the flow path or the calibrator pack that was replaced around the same time as the sensor in this case.It is difficult to confirm these findings as consumables were not returned.Therefore, no further actions are required at this time.Nova will continue monitoring for recurrence.
|