| Model Number |
67P045 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while preparing to do a routine tracheostomy tube change, the balloon was not inflating fully when checked prior to tube change.There were no visible signs of tears or leaks when balloon was inflated but "it didn't look right".Two trachs were observed to have this issue from the same lot number.A third trach was used without issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.The two used returned samples received and were inflated with air using a 5 ml syringe, after that the cuffs were gently manipulated.The whole cuff was inflated without abnormalities.It was demonstrated that the 2 cuffs were symmetrical, and no visible signs of tears or leaks were detected.Complaint is not confirmed.No root cause was determined due complaint was not confirmed.No corrective actions required since the complaint was not confirmed.D5 is unknown, no information provided to date.
|
| |
|