• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA FLEXTEND PEDIATRIC NEONATAL TRACHEOSTROMY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA FLEXTEND PEDIATRIC NEONATAL TRACHEOSTROMY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 67P045
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while preparing to do a routine tracheostomy tube change, the balloon was not inflating fully when checked prior to tube change.There were no visible signs of tears or leaks when balloon was inflated but "it didn't look right".Two trachs were observed to have this issue from the same lot number.A third trach was used without issue.
 
Additional Manufacturer Narrative
Other text: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.The two returned samples received were inflated with air using a 5 ml syringe, after that the cuffs were gently manipulated.The whole cuff was inflated without abnormalities.It was demonstrated that the 2 cuffs were symmetrical, and no leak was detected.Complaint is not confirmed.No root cause was determined due complaint was not confirmed; no corrective actions required.D5 is unknown, no information provided to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA FLEXTEND PEDIATRIC NEONATAL TRACHEOSTROMY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14945106
Report Number3012307300-2022-12961
Device Sequence Number4493986
Product Code JOH
UDI-Device Identifier15021312006285
UDI-Public15021312006285
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K944178
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/17/2025
Device Model Number67P045
Device Catalogue Number67P045
Device Lot Number3959301
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date07/07/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-