(b)(4).The investigation has been completed.Based upon the totality of the information received over the course of the investigation and reported by gore in the previously submitted medical record narratives the following conclusions have been reached.As previously reported, all pertinent medical records requested may not have been received.In the absence of additional information or medical records from the complainant, this event file will be closed with the information provided.Previous patient codes were reported based on the original complaint and are no longer applicable and/or not reportable per gore¿s investigation.It should be noted that the gore® dualmesh® biomaterial instructions for use include warnings and addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, contamination, infection, inflammation, adhesion, fistula formation, seroma formation, hematoma, and recurrence.Medical records that indicate mesh ¿movement¿ or that the device led to a recurrence may reflect a recurrence as a function of a patient¿s poor tissue quality leading to fascial dehiscence or loss of anchorage of fixation, or may be related to the hernia type, individual patient comorbidities, and technical and procedural aspects of the repair.These factors include but are not limited to, fixation type, mesh shape/sizing used, and defect closure decisions.Additionally, a new, unrelated hernia can occur but may be referred to as a recurrent hernia.As with any surgical procedure, there are always risks of complications for surgical repair of hernias and soft tissue deficiencies, with or without mesh.These may include but are not limited to, adhesions and related harms, bleeding, bowel obstruction, compromised device biocompatibility, contamination which may lead to patient harms, device damage, dysphagia, erosion or extrusion and related harms, exposure or protrusion and related harms, fever, fistula, gerd recurrence, defect recurrence and related harms, ileus, increased procedure time and related harms, irritation or inflammation, infection, mesh migration, mesh contraction, pain, paresthesia, perforation, revision/re-intervention, seroma or hematoma and related harms, tissue ischemia, wound complications and wound dehiscence and additional intervention including surgery.Many of the potential complications are associated with the patient¿s underlying disease progression, co-morbidities, additional medical history and/or other surgical procedures.The above inherent risks are typically detailed in standard informed consent documents.The device was not able to be returned to gore for evaluation; therefore, a direct product analysis could not be conducted.Review of the manufacturing records verified that the lot met all pre-release specifications.
|