• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24700
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the non-tortuous and mildly calcified cephalic vein.A 4.0mmx80mmx80cm sterling balloon catheter was advanced for dilatation.However, during the first inflation at 4 atmospheres, the balloon burst.The device was completely removed and the procedure was completed with another of the same device.There were no patient complications reported and patient status was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the non-tortuous and mildly calcified cephalic vein.A 4.0mmx80mmx80cm sterling balloon catheter was advanced for dilatation.However, during the first inflation at 4 atmospheres, the balloon burst.The device was completely removed and the procedure was completed with another of the same device.There were no patient complications reported and patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.The device showed signs of use.The sterling was examined and a rupture in the balloon was visible 7 mm form the tip with a length of 1.6 cm.No other damage was found on the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14947061
Report Number2124215-2022-24591
Device Sequence Number6215077
Product Code DQY
UDI-Device Identifier08714729123675
UDI-Public08714729123675
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K053116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/09/2025
Device Model Number24700
Device Catalogue Number24700
Device Lot Number0028843105
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/28/2022
Supplement Date Received by Manufacturer08/03/2022
Initial Report FDA Received Date07/07/2022
Supplement Report FDA Received Date08/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-