| Model Number |
24700 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the non-tortuous and mildly calcified cephalic vein.A 4.0mmx80mmx80cm sterling balloon catheter was advanced for dilatation.However, during the first inflation at 4 atmospheres, the balloon burst.The device was completely removed and the procedure was completed with another of the same device.There were no patient complications reported and patient status was stable.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the non-tortuous and mildly calcified cephalic vein.A 4.0mmx80mmx80cm sterling balloon catheter was advanced for dilatation.However, during the first inflation at 4 atmospheres, the balloon burst.The device was completely removed and the procedure was completed with another of the same device.There were no patient complications reported and patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.The device showed signs of use.The sterling was examined and a rupture in the balloon was visible 7 mm form the tip with a length of 1.6 cm.No other damage was found on the device.
|
| |
|