| Model Number |
TSM-330 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, an opening in the posterior side assessed, as unidentified tear opening and (the thickness within specification).Additional observations: wear abrasion observed, in the device.Crease fold observed, in the device.
|
| |
|
Search Alerts/Recalls
|