Model Number SVMB13
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant was removed because it was fractured.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).Additional pma/510(k) number ¿ k011028 | k013227.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).The reported device was returned for investigation.Visual inspection of the returned product identified what appeared to be bone / tissue attached to the implant external threads and a fracture at the collar of the implant no pre-existing conditions were noted on the product experience report (per).The reported device was located on tooth #23 and was used for approximately 5 years.Complainant reported that the implant was removed because it was fractured.The reported complaint was confirmed.A definitive root cause for this complaint could not be determined.A device history record (dhr)review was performed and no related deviations or nonconformances were noted.Also, a complaint history search was performed using our complaint handling system and there were no related complaints for this product lot.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL SCR-V MINI SBM 3.3MM 3.5MM 13MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14948412
Report Number0002023141-2022-01740
Device Sequence Number12299100
Product Code DZE
UDI-Device Identifier00889024018136
UDI-Public(01)00889024018136(17)220331(10)63582758
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K061410
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/31/2022
Device Model NumberSVMB13
Device Catalogue NumberSVMB13
Device Lot Number63582758
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date07/07/2022
Supplement Report FDA Received Date01/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/09/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-