| Model Number |
SVMB13 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implant was removed because it was fractured.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number (b)(4).Additional pma/510(k) number ¿ k011028 | k013227.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number (b)(4).The reported device was returned for investigation.Visual inspection of the returned product identified what appeared to be bone / tissue attached to the implant external threads and a fracture at the collar of the implant no pre-existing conditions were noted on the product experience report (per).The reported device was located on tooth #23 and was used for approximately 5 years.Complainant reported that the implant was removed because it was fractured.The reported complaint was confirmed.A definitive root cause for this complaint could not be determined.A device history record (dhr)review was performed and no related deviations or nonconformances were noted.Also, a complaint history search was performed using our complaint handling system and there were no related complaints for this product lot.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|