This report is being supplemented to provide additional information based on the approved final investigation and device evaluation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it was not possible to inject liquid when the slider was pushed.However, it was possible to inject liquid when the slider was pulled.The outer tube was not buckled, and the needle tube presented compressive buckling.However, the definitive root cause of the injector malfunction could not be determined.The likely root cause of the injector malfunction is compressive buckling of the needle tube.Since friction resistance between the outer tube and needle tube has increased, compressive buckling might have occurred when the needle was extended.Friction resistance between the outer tube and needle tube might have been cause by the tube having been coiled during the inspection for operation, or the slider having been pushed abruptly.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ·straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.Operate the slider slowly, otherwise the tube could buckle.Do not coil the insertion portion with a diameter of less than 15 cm.This could damage the insertion portion.When inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.Insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid." olympus will continue to monitor field performance for this device.
|