• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. MYCARELINK PATIENT MONITOR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. MYCARELINK PATIENT MONITOR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number 24950M
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the remote monitor started to burn after plugging it.It was further reported that the remote monitor was too hot to touch.No troubleshooting was taken.The monitor is expected to be replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product analysis: model: 24950m, serial/lot#: (b)(6): the remote monitor was returned and analysis revealed a radio frequency (rf) head battery failure.In addition, the remote monitor was stuck at error code 7332 during boot up.Moreover, the rf head battery was found defective during functional test.Lastly, the unit was charged for 30 minutes.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The remote monitor was replaced and returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYCARELINK PATIENT MONITOR
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14950003
Report Number2182208-2022-02232
Device Sequence Number6195710
Product Code NVZ
UDI-Device Identifier00763000367800
UDI-Public00763000367800
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number24950M
Device Catalogue Number24950M
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2022
Supplement Date Received by Manufacturer10/20/2022
Initial Report FDA Received Date07/07/2022
Supplement Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
-
-