| Model Number |
MFS141 |
| Medical Device Problem Code |
Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that several bd maxguard flow controller extension set with needleless y-site(s) experienced issue with the regulators.The following information was provided by the initial reporter: we have been receiving complaints from multiple nurses stating that the iv set regulators (dial a flow) are malfunctioning.The regulators are working properly when they prime, flush the line, and turn the drip off.Once they reset the dial to the appropriate drip rate, after turning it off, the intermittent issue of the regulator not flowing occurs.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Multiple lot numbers: there were multiple lot numbers reported to be involved.The information for additional lot number is as follows: medical device lot #: 22039076.Medical device expiration date: 2025-03-04.Device manufacture date: 2022-03-04.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: two sample for model mfs141 and lot number# 22039076 was returned for investigation.One was used and another one was unused.The set was examined for defects and abnormalities.No defects or abnormalities were observed with unused set, but observed with used set.The used set was attached to a bd syringe to see if the issue is with the regulator.The dial was set to the appropriate drip rate as open, 200, 100 and 50 drip position, and a fluid path was observed, the intermittent issue of the regulator (dial a flow) occurs.The customer complaint about the extension set with the flow issues could be replicated.With the unused set the regulator is working properly when prime, flush the line, and turn the drip for different rate as open , 200 ,100 and 50 drip position.An investigation had been conducted at the supplier and corrective actions have been implemented in order to correct the issue.The product in question was created before changes to the manufacturing process were implemented.No sample was returned for model mfs141 and lot number# 22039098 so no investigation performed on it.The root cause is caused by a misplaced gasket.Corrective actions have been taken to correct the issue and the component that was received was constructed before the corrective actions were put in place.The root cause of intermittent issue of the regulator not flowing occurs could not be determined for unused sample because the issue could not be replicated.A device history record review for model mfs141 and lot number 22039076 was performed.The search showed that a total of 14,403 units in 1 lot number was built on 07mar2022.A device history record review for model mfs141 and lot number 22039098 was performed.The search showed that a total of 14,403 units in 1 lot number was built on 08mar2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set for the lots # 22039076 and #22039098.
|
| |
|
Event or Problem Description
|
|
It was reported that several bd maxguard flow controller extension set with needleless y-site(s) experienced issue with the regulators.The following information was provided by the initial reporter: we have been receiving complaints from multiple nurses stating that the iv set regulators (dial a flow) are malfunctioning.The regulators are working properly when they prime, flush the line, and turn the drip off.Once they reset the dial to the appropriate drip rate, after turning it off, the intermittent issue of the regulator not flowing occurs.
|
| |
|