Model Number PRD-SC30-MICRO
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 06/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During treatment in the distal anterior tibial artery and dorsalis pedis arteries, the stealth 360 peripheral orbital atherectomy device (oad) stalled while in the mid dorsalis pedis artery.The patient experiences restless leg syndrome and when intravascular ultrasound was employed, the vessel spasmed.When the oad was removed, a medial and inferior perforation of the dorsalis pedis artery was observed.Angioplasty was performed and left in place for 2 minutes to tamponade the vessel.Final angioplasty revealed a perforation was still present.The procedure was completed and, compression dressing was placed on access site on foot.
 
Additional Manufacturer Narrative
The oad was received at csi for analysis.Review of the oad's data log, confirmed the reported stall.The root cause of the stall events was not determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The diamondback 360® peripheral orbital atherectomy system instructions for use manual states that perforation and vessel spasm are potential adverse events that may occur and/or require intervention with use of the system.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
kim mcclure
1225 old hwy 8 nw
st. paul, MN 55112
6519006741
MDR Report Key14950106
Report Number3004742232-2022-00157
Device Sequence Number6148238
Product Code MCW
UDI-Device Identifier10850000491264
UDI-Public(01)10850000491264(17)230831(10)393313-1
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberPRD-SC30-MICRO
Device Catalogue Number7-10059-01
Device Lot Number393313-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/07/2022
Initial Report FDA Received Date07/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexFemale
Patient RaceWhite
-
-