MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS ESHEATH INTRODUCER SET; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS ESHEATH INTRODUCER SET; INTRODUCER, CATHETER Back to Search Results
Model Number 914ESA
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Vascular Dissection (3160); Thrombosis/Thrombus (4440)
Date of Event 09/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Per the instructions for use (ifu), cardiovascular injuries such as perforation or dissection of vessels, ventricle, myocardial or valvular structures are known potential risks or adverse events associated with the overall thv procedure and may require intervention.According to the literature review, and as documented in a technical summary written by edwards lifesciences, vascular complications are a well-recognized complication of the transfemoral thv procedure in this elderly population with multiple co-morbidities.Edwards has reviewed many reports, including screening data records and source documentation of vascular complications, and has found that the root cause is typically related to a combination of vessel size, tortuosity, and calcifications.Although the incidence is decreasing with smaller sheath/delivery system sizes and physician experience, there will continue to be cases in which vascular complications will occur.The thv physician training manuals instruct on procedural considerations for sheath insertion with regards to proper screening critical to reducing vascular complications.The training manual instructs the operator on proper sheath insertion and withdrawal techniques, including pre-dilating the vessel with the edwards dilators, as needed.It also notes that calcification may reduce lumen diameter and limit or prevent the transfemoral passage of the devices.The ifu contraindicates patients with ilio-femoral vessel characteristics that would preclude safe placement of sheaths such as severe obstructive calcification or severe tortuosity.Pre-procedure screening and assessment of the femoral/iliac artery internal diameters will enable the clinician to determine if the sapien valve can be delivered transfemorally.Assessment of location and amount of circumferential calcium will aid in determining areas of reduced vessel diameters.The operators are trained to measure minimum vessel diameter taking calcium into account.The physician training manual also lists the minimum recommended vessel size for each size device.Despite the best screening tools, a small percentage of patients will have femoral/iliac vessels that are not amenable to the trans-femoral approach or where increased resistance is encountered during insertion of devices.In many cases, the vessel minimum luminal diameter (mld) may be borderline or below the indicated size.In addition, significant calcification and/or tortuosity, not always appreciable on imaging, could be contributing factors to the event.The minimum required vessel diameter for an 14fr esheath is 5.5 mm.In this case, there was no allegation or indication a device malfunction contributed to this adverse event.Per the instructions for use (ifu), potential risks associated with the overall procedure include thrombus formation, plaque dislodgment, and embolization that may result in myocardial infarction, stroke, distal peripheral occlusion, and/or death.Per the instructions for use (ifu), vascular dissection and thrombus are potential adverse events associated with the transcatheter valve replacement (tvr) procedure and the use of the edwards thv devices.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest procedural factors may have- caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
 
Event or Problem Description
As reported through social media, a dissection of the femoral artery occurred during the tavr procedure.Two days post discharge, the patient returned to the hospital due to a blood clot in the femoral artery that required surgical intervention.As reported by a tavr patient through social media, during a transfemoral tavr procedure, during the attempt to place the sheath, the right femoral artery was dissected.No actions were taken as the doctor believed the dissection would heal on its own.A 26mm sapien 3 ultra valve was successfully deployed in the aortic position.Two days post discharge, 'tightness' in the right calf occurred and the patient returned to the hospital.Surgery was performed to remove a blood clot at the access site.The patient was discharged 4 days post-surgery.The patient reported a quick recovery and feeling well.The valve remains implanted in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS ESHEATH INTRODUCER SET
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
MDR Report Key14950428
Report Number2015691-2022-06657
Device Sequence Number20180824
Product Code DYB
UDI-Device Identifier00690103193985
UDI-Public(01)00690103193985(17)220415(10)63130340
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K200258
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/15/2022
Device Model Number914ESA
Device Lot Number63130340
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2022
Initial Report FDA Received Date07/07/2022
Date Device Manufactured04/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-