Model Number CRYOS
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2021 when preparing for a modified pectus excavatum procedure, the cryos probe was connected to the acm generator, laid on a tabletop and a test freeze was initiated.When the probe reached -40 degrees, the acm counted down for about 10 seconds and then scrub technician vented it by pressing the n2o manual exhaust knob.When doing this a loud "pop" sound occurred and the device handle separated.There was no patient harm.During device evaluation, it was observed that the exhaust crimp sleeve was not present in the device.This discovery escalated this product problem to a reportable event per part 803 on (b)(6) 2022.
 
Additional Manufacturer Narrative
(b)(4) the complaint was confirmed.  the internal crimp sleeve component was not properly assembled to the exhaust tube.  the exhaust tube slid loose from the mating barb which caused the loud noise.  atricure is conducting an investigation to better understand the circumstances involving the improperly assembled crimp sleeve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYOICE CRYOSPHERE CRYO-ABLATION PROBE
Common Device Name
CRYOICE CRYOSPHERE CRYO-ABLATION PROBE
Manufacturer (Section D)
ATRICURE, INC.
7555 innovation way
mason OH 45040
Manufacturer (Section G)
ATRICURE, INC.
7555 innovation way
mason OH 45040
Manufacturer Contact
john ehlert
7555 innovation way
mason, OH 45040
5136448220
MDR Report Key14950614
Report Number3011706110-2022-00024
Device Sequence Number20559175
Product Code GXH
UDI-Device Identifier00818354016321
UDI-Public00818354016321
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K182565
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2024
Device Model NumberCRYOS
Device Catalogue NumberA001009
Device Lot Number113029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2022
Initial Report FDA Received Date07/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACM: 44569-05
Patient SexUnknown
-
-