Model Number VCP218H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Photo analysis: this is an analysis of a photo submitted for evaluation.During the visual analysis the following was observed: the photograph shows a winding former, needle/suture needle tip not present because it was poorly formed.The missing needle tip defect has been correlated to the manufacturing process.Based on currently available information, the missing needle tip was identified during investigation of the sample received.The problem of this product will be addressed through the ethicon quality system.Additional monitoring of complaint information for potential safety signals will be conducted through complaint trends as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name - irgacare.Active ingredient(s)- triclosan.Dosage form - suture/solid/parenteral.Strength -= 275 g /m.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and suture was used.During use on the patient, it was reported that there had been no needle tip on one needle in the newly dispensed suture when it was opened to prepare for surgery.Changed another like device to complete the surgery.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 8/26/2022.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigational summary: the product received for analysis was identified as product code vcp218h.Visual inspection and metallurgical analysis were performed on the returned product.A fracture was observed a blunt tip.A scanning electron microscope was used to examine the fracture surfaces and surrounding area of the needle.The surfaces were examined in multiple locations to determine the cause.The evaluation revealed the needle contained a blunt tip.The area with the blunt tip contained mechanical damage and some ductile tearing.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Trade name - irgacare; active ingredient(s)- triclosan; dosage form - suture/solid/parenteral; strength -= 275 g /m.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 27IN 4-0 S/A SH-1
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN ANGELO
3348 pulliam st
san angelo TX 76905 4403
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key14951664
Report Number2210968-2022-05233
Device Sequence Number5485742
Product Code GAM
UDI-Device Identifier10705031051706
UDI-Public10705031051706
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K132580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberVCP218H
Device Catalogue NumberVCP218H
Device Lot NumberRJ2ACP
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2022
Supplement Date Received by Manufacturer08/16/2022
Initial Report FDA Received Date07/07/2022
Supplement Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-