| Model Number |
VCP218H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Photo analysis: this is an analysis of a photo submitted for evaluation.During the visual analysis the following was observed: the photograph shows a winding former, needle/suture needle tip not present because it was poorly formed.The missing needle tip defect has been correlated to the manufacturing process.Based on currently available information, the missing needle tip was identified during investigation of the sample received.The problem of this product will be addressed through the ethicon quality system.Additional monitoring of complaint information for potential safety signals will be conducted through complaint trends as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name - irgacare.Active ingredient(s)- triclosan.Dosage form - suture/solid/parenteral.Strength -= 275 g /m.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and suture was used.During use on the patient, it was reported that there had been no needle tip on one needle in the newly dispensed suture when it was opened to prepare for surgery.Changed another like device to complete the surgery.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 8/26/2022.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigational summary: the product received for analysis was identified as product code vcp218h.Visual inspection and metallurgical analysis were performed on the returned product.A fracture was observed a blunt tip.A scanning electron microscope was used to examine the fracture surfaces and surrounding area of the needle.The surfaces were examined in multiple locations to determine the cause.The evaluation revealed the needle contained a blunt tip.The area with the blunt tip contained mechanical damage and some ductile tearing.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Trade name - irgacare; active ingredient(s)- triclosan; dosage form - suture/solid/parenteral; strength -= 275 g /m.
|
| |
|
Search Alerts/Recalls
|
|
|