• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P TAPERFILL HIP; TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 7

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P TAPERFILL HIP; TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 7 Back to Search Results
Model Number 425-97-007
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 06/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as stem subsided.The previous surgery and the surgery detailed in this event occurred 5 months and 2 weeks apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to stem subsided.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery: the initial stem subsided surgeon removed the 7 taperfill stem and used a stryker revision stem to replace it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERFILL HIP
Common Device Name
TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 7
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key14951973
Report Number1644408-2022-00901
Device Sequence Number6896274
Product Code LPH
UDI-Device Identifier00888912078658
UDI-Public(01)00888912078658
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130099
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number425-97-007
Device Catalogue Number425-97-007
Device Lot Number172R1211
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2022
Initial Report FDA Received Date07/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
942-01-36H LOT 624Z1047
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age72 YR
Patient SexMale
-
-