Model Number 8637-40
Device Problems Electromagnetic Interference (1194); Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/22/2022
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from an unknown source regarding a patient who was receiving morphine (10 mg/ml at 2.527 mg/day) via an implantable pump for non-malignant pain.It was reported that the patient was in yesterday and the healthcare provider (hcp) noted a pump motor stall occurred on (b)(6) 2022.The patient has acute and long-term memory issues and they were unsure if they had magnetic resonance imaging (mri) on that date.They did hear the pump alarm after it was interrogated yesterday and saw a motor stall recovery that occurred yesterday as well.The hcp put the pump in minimum rate mode and planned to bring the patient back in again.Technical services reviewed information around telemetry state and advised caller to try and confirm whether or not the patient had an mri on (b)(6) 2022.Advised hcp to check pump reservoir to measure volume for accuracy.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the healthcare provider (hcp) confirmed that the cause of the motor stall was due to an mri in april.The report showed "stopped pump > 48" on (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14951997
MDR Text Key295500122
Report Number3004209178-2022-08603
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169530126
UDI-Public00643169530126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2019
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/28/2022
Initial Date FDA Received07/07/2022
Supplement Dates Manufacturer Received08/24/2022
Supplement Dates FDA Received08/26/2022
Date Device Manufactured04/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexFemale
Patient Weight89 KG
-
-