• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G52918
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
Description of event according to initial reporter: when the physician tried to deploy the filter, it would not come off of the jugular deployment system.The filter would not come off after pushing the release.A new g52918-igtcfs-65-1-uni-tulip was opened and used to complete the intended procedure successfully.Patient outcome: the patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Manufacturer ref# (b)(4).Initial reporter occupation: director of radiology.Pma/510(k): k211874.Investigation is still in progress.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Manufacturers ref# (b)(4).After investigation the event for (b)(6) is no longer reportable.Summary of investigational findings: when the physician tried to deploy the filter, it would not come off the deployment system from jugular approach.A new device was used instead to complete the intended procedure successfully.It was assessed that because any discovered non-conformances were properly dispositioned before qc release, there is evidence that the dhr was fully executed.In addition, no other lot related complaints have been received from the field.It is therefore concluded that there is no evidence that nonconforming product exists in house or in field.According to the instruction for use excessive tension during deployment may prevent the filter from releasing when the release mechanism is activated.There are adequate controls in place to ensure the device was manufactured to specifications.No device or images were provided for the investigation and without the actual device it would be inappropriate to speculate what may or may not have caused the deployment difficulties.However, excessive tension during deployment could have prevented the filter from being released but this is purely speculation.This report is required by the fda under 21cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET
Common Device Name
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
MDR Report Key14954272
Report Number3002808486-2022-00884
Device Sequence Number6915486
Product Code DTK
UDI-Device Identifier10827002529189
UDI-Public(01)10827002529189(17)241118(10)E4187491
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/18/2024
Device Model NumberG52918
Device Catalogue NumberIGTCFS-65-1-UNI-TULIP
Device Lot NumberE4187491
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/01/2022
Supplement Date Received by Manufacturer08/05/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date08/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-