| Catalog Number |
627336403 |
| Medical Device Problem Code |
Device Misassembled During Manufacturing /Shipping (2912)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was not provided.Sorin group italia manufactures the smart perfusion pack.The incident occurred in (b)(6).If any additional information pertinent to the reported event is obtained, it will be provided in a supplemental report.Device not available.
|
| |
|
Event or Problem Description
|
|
Livanova usa received a report that during installation of the circuit, the tubing disconnected from the connector even when its tie banded.There is no report of any patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Livanova received a report stating that during priming the tubing just slipped back off the 3/8 connector even when it was tie banded.Picture of the issue was received, confirming it.Pack was not available for return.The dhr review highlighted that the lot whose claimed product belong to was released as conform according to specifications.By reviewing the complaints database, a further similar event for tubing disconnection related to this connector has been recorded on the same pts pack code, but different lot number.This involved 3/8 connector is assembled with the tubing through the use of solvent and then tie banded.Based on above and without sample to analyze, it cannot be ruled out that the disconnection was due to the following possible root causes: - insufficient presence of solvent, - tie band not tight enough, - tubing and/or connector dimensional issue.The risk is in the acceptable region.Livanova will keep monitoring the market.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|
Search Alerts/Recalls
|