It was reported to boston scientific corporation that a contour vl ureteral stent was used during a removal of kidney stones procedure in the right kidney performed on (b)(6) 2022.During the procedure, when the stent was introduced and passed through the urethra and entered the forced of two centimeters into the ureter, and when the stent was pushed, it became buckled and would not advance into the ureter and a torn was observed on the distal part.The procedure was successfully completed with another contour vl ureteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.No further information has been obtained despite good faith efforts.
|
|
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
|
It was reported to boston scientific corporation that a contour vl ureteral stent was used during a removal of kidney stones procedure in the right kidney performed on (b)(6) 2022.During the procedure, when the stent was introduced and passed through the urethra and entered the forced of two centimeters into the ureter, and when the stent was pushed, it became buckled and would not advance into the ureter and a torn was observed on the distal part.The procedure was successfully completed with another contour vl ureteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.No further information has been obtained despite good faith efforts.
|
|
Block h6: medical device code a040601 captures the reportable event of stent buckled, inside the patient.Medical device code a0414 captures the reportable event of stent torn, inside the patient.Block h10: the returned contour vl ureteral stent was analyzed, and a visual evaluation noted that with three unopened pouches, the positioner, guide wire and axxcess catheter.Their stabilizer was also returned in good conditions.Then found that the bladder pig tail was buckled accordion, side port holes stretch out, suture hole and bladder coil tip torn.Additionally, their suture was not returned.No other problem with the device were noted.The reported event was confirmed.Based on the most relevant information of the complaint event description, device analysis and including the product record review results, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.It was found that the device has the bladder coil buckled or accordion.Also, during magnification, it was found that side port holes were stretched, their suture hole and tip torn.Additionally, the suture was not returned, which means that it was removed, indicating that the stent was used.According to the evidence, it is possible that operational factors, such as interacting of the device with positioner, interaction with another device could cause the buckled or accordion and the holes got stretch and the tip torn.This issue could at the same time cause that the device being difficult to advance in anatomy, consequently, affect the performance of the device.Additionally, an interaction of the device with their suture such as an entanglement could have contributed to the suture hole torn.Therefore, the most probable cause of this complaint is adverse event related to procedure since it is the most likely that the adverse event occurred during the procedure and the device had no influence on event.
|