MAUDE Adverse Event Report: COOK INC OPEN-END FLEXI-TIP URETERAL CATHETER; KOD CATHETER, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC OPEN-END FLEXI-TIP URETERAL CATHETER; KOD CATHETER, UROLOGICAL Back to Search Results
Model Number G14521
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, the tip of a open-end flexi-tip ureteral catheter broke off inside of the patient's body.The tip was retrieved intra-operatively.Additional event and patient outcome information has been requested.
 
Additional Manufacturer Narrative
Device return is unknown.Initial reporter occupation: coordinator.Device evaluated by mfg.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown or unavailable.Event description: as reported, the tip of a open-end flexi-tip ureteral catheter broke off inside of the patient's body.The tip was retrieved intra-operatively.Investigation - evaluation reviews of complaint history, device history record (dhr), instructions for use (ifu) and quality control procedures were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed; however, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record found no non-conformances related to the reported failure mode.The pull test report in the work order was reviewed and the lot was confirmed to have passed pull testing.A review of complaint history records shows no other related complaints associated with the complaint device lot.Because there were no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in the field.A review of relevant manufacturing and quality control documents was conducted.Cook has concluded that controls are in place to assure functionality and device integrity prior to shipping.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: precautions avoid bending or kinking the catheter prior to placement.Doing so could damage the integrity of the catheter and result in patient injury.If you encounter resistance while advancing or withdrawing the catheter, stop.Determine the cause of the resistance before proceeding.Do not withdraw catheter while deflected in cystoscope/endoscope.A cause for the complaint could not be established.It was not possible to rule out patient factors/condition, [inadvertent] user/procedural issues and concomitant device use.Per the quality engineering risk assessment, no further action is required.The appropriate personnel have been notified, and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN-END FLEXI-TIP URETERAL CATHETER
Common Device Name
KOD CATHETER, UROLOGICAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14955775
Report Number1820334-2022-01171
Device Sequence Number5396378
Product Code KOD
UDI-Device Identifier00827002145214
UDI-Public(01)00827002145214(17)250406(10)14650652
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K171662
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/06/2025
Device Model NumberG14521
Device Catalogue Number021305
Device Lot Number14650652
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/23/2022
Supplement Date Received by Manufacturer08/18/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date09/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-