MAUDE Adverse Event Report: ZIMMER DENTAL IMP,TM 3.7MM MTX,10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP,TM 3.7MM MTX,10MM; DENTAL IMPLANT Back to Search Results
Model Number TMMB10
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Doctor reported that an implant fracture at tooth site #32.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4) the following sections have been updated: b4: date of this report b5: describe event or problem g3: date received by the manufacturer h1: type of report, follow-up number h2: follow up type h3: device evaluated by manufacturer: change ¿no' to 'yes' h6: evaluation codes h10: additional narrative one implant was returned for investigation.Visual evaluation of the as returned implant identified implant fractured at the collar region.Signs of use.Functional testing could not be performed since the product was fractured.Pre-existing conditions noted on the per were 'low bone density ¿ type iii'.The reported device had been placed on tooth 32 (fdi) for approximately 5 months.The patient is a smoker x-ray & picture evaluation: picture or x-ray evidence not provided.Review of appropriate documentation: documents reviewed: instructions for use - tapered screw-vent and trabecular metal implants, ifu4869 rev 9-10/19.Information identified: contraindications & precautions.Per the applicable ifu, it is stated that improper techniques, severe bruxism, clenching, and overloading, may cause component fracture.Dhr review: dhr review for the lot (1232699) had revealed no deviations nor non-conformances which could have caused or contributed to the reported events.All products were conforming at the time they left zimvie.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: complaint history review was performed for the reported lot number (1232699) for similar events (complaint category keywords: fracture: implant) and no other complaint was identified.Post market trending review: october post market trending was reviewed and there were no actionable events or corrective actions for the reported event or product.Therefore, based on the available information, device malfunction did occur, and the reported event was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TM 3.7MM MTX,10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14956168
Report Number0002023141-2022-01755
Device Sequence Number15191809
Product Code DZE
UDI-Device Identifier00889024018860
UDI-Public(01)00889024018860(17)241013(10)1232699
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K132258
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/13/2024
Device Model NumberTMMB10
Device Catalogue NumberTMMB10
Device Lot Number1232699
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
-
-