| Model Number |
72202971 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case- (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified procedure, when drilling the cortex of the femur with the 4.5mm flexible drill, it split and it had to be drilled with the auger 5.0mm from the outside to the inside.It is unknown how the procedure was completed or if there was a surgical delay.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: b5 updated.
|
| |
|
Event or Problem Description
|
|
It was reported that during an acl reconstruction, lateral condile procedure, when drilling the cortex of the femur with the 4.5mm flexible drill, it split and it had to be drilled with the auger 5.0mm from the outside to the inside.The procedure was completed with a 3 minutes surgical delay using a back-up device in the originally drilled bone hole.No further complications were reported.The current status of the patient is well.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).H10 h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the customer supplied images finds partial images of the complaint device.It appears the blade portion is fractured away.This case reports, the flexible drill split during drilling in an acl reconstruction, lateral condyle procedure and the auger 5.0 mm was then drilled from outside to inside.Per case details, the patient¿s current health status is well.However, it is unknown if the drill broke internal to the patient or if any broken pieces fell inside the patient.The material composition of the of the flexible drill is 465 sst.No clinically relevant supporting documentation was provided for inclusion in this medical investigation.The drills are manufactured and intended as externally communicating devices and are not approved for long term internal tissue exposure and long-term implantation data is not available.The patient impact beyond possible micro-motion and/or migration, and local irritation/discomfort cannot be determined.No further medical assessment can be rendered at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the reported device was returned to the designated complaint unit for independent evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.There were 3 devices returned.A visual inspection of the complaint device found that it is not in its original packaging.The drill is warped through the area of flexibility and the distal tip is broken away on an unknown guide wire.There is biological debris in the device.A visual inspection of an additional returned device found that it is not in its original packaging.The lot number is 50821659 drill is warped through the area of flexibility and the proximal end of the flexible portion is cracked.There is biological debris in the device.A visual inspection of an additional returned device found that it is not in its original packaging.The lot number is 50864013 drill is warped through the area of flexibility and the proximal end of the flexible portion is bent at almost 90 deg.There is biological debris in the device.A review of the customer supplied images finds partial images of the complaint device.It appears the blade portion is fractured away.¿¿a review of the customer supplied images finds partial images of the complaint device.It appears the blade portion is fractured away.This case reports, the flexible drill split during drilling in an acl reconstruction, lateral condyle procedure and the auger 5.0 mm was then drilled from outside to inside.Per case details, the patient¿s current health status is well.However, it is unknown if the drill broke internal to the patient or if any broken pieces fell inside the patient.The material composition of the of the flexible drill is 465 sst.No clinically relevant supporting documentation was provided for inclusion in this medical investigation.The drills are manufactured and intended as externally communicating devices and are not approved for long term internal tissue exposure and long-term implantation data is not available.The patient impact beyond possible micro-motion and/or migration, and local irritation/discomfort cannot be determined.No further medical assessment can be rendered at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|